PRODUCT OVERVIEW
5
Water Specifications
Purified water specification Crystal EX Trace Crystal EX HPLC Crystal EX Bio
Grade II resistivity at 25 °C
>10 M x cm >10 M x cm >10 M x cm
Grade II conductivity at 25 °C <0.1 µS/cm <0.1 µS/cm <0.1 µS/cm
Grade I resistivity at 25 °C
18.2 M x cm 18.2 M x cm 18.2 M x cm
Grade I conductivity at 25 °C 0.055 µS/cm 0.055 µS/cm 0.055 µS/cm
TOC level <10 ppb <2 ppb <2 ppb
RNase N/A N/A <0.01 ng/mL
DNase N/A N/A <4 pg/
L
Bacteria <0.01 CFU/mL <0.01 CFU/mL <0.01 CFU/mL
Endotoxins <0.15 EU/mL <0.15 EU/mL <0.001 EU/mL
Technical Specifications
Dimensions and weight
Crystal EX Trace Crystal EX HPLC Crystal EX Bio
Dimensions (W*D*H) 40*35*55 cm 40*35*55 cm 40*35*55 cm
System weight 17 kg 18 kg 19 kg
Operating weight 24 kg 25 kg 26 kg
Noise level
Water purification system Crystal EX can generate a maximum sound pressure level of 47 dB at 1 m
distance from the system.
Electrical requirements
The water purification system is configured for 230 V ± 5%, 600 mA max.
In areas where the supplied power is subject to voltage fluctuations exceeding 10% of the nominal
volume, a power line regulator may be required.
Power supply receptacle should be within 1.5 meters from the installation site.