UNPACKING
9
Site requirements
The system requires up to 400*350*550 mm (W*D*H) space on the bench. If the system is equipped
with the tank, allow enough space for the tank. The tank can be placed under the bench.
Environmental requirements
The water purification system is intended for indoor use only, in an environment that has
nonconductive pollutants only.
Ensure that the site is maintained under the following conditions:
Condition Acceptable range
Temperature 15 to 30 °C (59 to 86 °F)
Humidity 20% to 80% relative humidity, noncondensing
Water leakage safety
Make sure that all water connection tubes are kink-free.
Make sure that all water connections are tight.
When installing the tank, connect the fitting marked “OVERFLOW” to the drain. Make sure
that the drain level is lower than the “OVERFLOW” fitting level. This will prevent water
leakage in case of tank level sensor failure.
A 1 um sediment filter has to be installed in the feedwater supply line. Failure to install the
filter may result in clogging of the strainer inside the system and water flow blockage.
Additional equipment
Water purification systems Crystal EX Trace, Crystal EX HPLC and Crystal EX Bio are not equipped with
internal tank.
Therefore, you may need to order external tank for purified water storage additionally. See Price list
for available options.
UNPACKING
Remove packing materials carefully and retain for them future shipment or storage of the unit. Examine
the unit carefully for any damage incurred during transit. The warranty does not cover in-transit
damage.
Due to the unit’s weight its unpacking and installing is to be carried out by two persons.
When opening the shipping box, we advise to compare received parts with the Packing List included.
Contact Adrona if any part is missing.