MAINTENANCE
33
Cleaning and Sanitization of the system
Clean the exterior of the system with soft tissue and water. Do not use any chemicals.
The water storage tank should be cleaned and disinfected every 2-3 months.
Prior to cleaning, you need to disconnect the water tank from the system. Place JG plugs in
all tube connectors.
Remove storage tank cap and add hydrogen peroxide to the tank. Final solution should be 1% hydrogen
peroxide.
30% hydrogen peroxide solution can damage polypropylene components.
Use safety equipment when handling unsafe liquids.