MAINTENANCE
40
Empty the tank
[1]
[2]
1 – Open the water dispense tap;
2 – Empty the tank;
3 – When the water does not flow
anymore, open the “REC OUT” tube;
4 – Pour out the remaining water
Adrona Recirculation Sanitization
1 – Allow for Adrona water purification system to fill the tank.
2 – It is possible to proceed with sanitization with partial tank by disconnecting tank level sensor.
3 – Press “Run” button to switch unit to “StandBy” mode.
4 – Disconnect pre-filter and Polishing module. Connect pre-filter and Polishing module outlets
with a bypass (see picture below for reference):
Front panel Rear panel
5 – Remove storage tank cap and add hydrogen peroxide to the tank. Final solution should be 1%
hydrogen peroxide.
30% hydrogen peroxide solution can damage polypropylene components.
Use safety equipment when handling unsafe liquids.
6 – After adding sanitization liquid switch unit to running mode.
7 – It is advised to leave recirculation interval and recirculation time to the standard 10 and 5
minutes respectively.