PRODUCT OVERVIEW
5
Water Specifications
Purified water
specification Q-Front N Trace Q-Front N HPLC Q-Front N Bio Q-Front N Bio UF
Grade I water resistivity
Grade I water
conductivity
0.055 μS/ cm 0.055 μS/ cm 0.055 μS/ cm 0.055 μS/ cm
Grade II water
conductivity
<0.1 μS/cm <0.1 μS/cm <0.1 μS/cm <0.1 μS/cm
TOC <10 ppb <2 ppb <2 ppb <2 ppb
DNase - - <4 pg/μL -
Bacteria <0.01 CFU/mL <0.01 CFU/mL <0.01 CFU/mL <0.01 CFU/mL
Endotoxins <0.15 EU/mL <0.15 EU/mL < 0.001 EU/mL -
Particles >0.22 μm <1/ per mL <1/ per mL - -
Nominal flow, pure
Dispense rate, ultrapure
water Up to 2 L/min Up to 2 L/min Up to 2 L/min Up to 2 L/min
Technical Specifications
Dimensions and weight
Dimensions (W*D*H) 35*39*54 cm
Operating weight 30 kg 31 kg 32 kg
Noise level
Water purification system Q-Front N can generate a maximum sound pressure level of 47 dB at 1 m
distance from the system.
Electrical requirements
The water purification system is configured for 230 V ± 5%, 3.15 A max.
In areas where the supplied power is subject to voltage fluctuations exceeding 10% of the nominal
volume, a power line regulator may be required.
Power supply receptacle should be within 1.5 meters from the installation site.