EasyManua.ls Logo

Advanced Instruments 3320 - 3 Calibration; Understanding calibration; Calibration procedure

Advanced Instruments 3320
92 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3
Calibration
Understanding calibration
It must be emphasized that repeatability is essential to calibra-
tion. A "calibration" is based on repeatable "hard" numbers for
reference materials. Reference values can be plotted on a
graph, or the instrument can be adjusted to be direct reading.
Direct-reading calibration of the Advanced Instruments Model
3320 is a simple procedure that requires no adjustment of the
instrument by the user. Simply run the menu-driven calibration
program, which prompts the user to run a series of tests using
calibration standards at 50 mOsm/kg H
2
O, 850 mOsm/kg H
2
O,
and if selected, 2000 mOsm/kg H
2
O. When the tests are com-
plete, the Advanced Instruments Model 3320 then calculates and
stores the calibration coefficients.
To optimize the performance of the Advanced Instruments
Model 3320, an optional third calibration point at 2000
mOsm/kg H
2
O can be used. Please refer to Chapter 4 - User
Interface: Setup Menu for more information.
Calibration procedure
1. When the display reads "Osmometer Ready", press the
[NEXT] button until [CALIB] appears over the left button.
Press it to initiate the calibration procedure. Calibration
can be cancelled without changing the existing calibration
by pressing the [EXIT] button.
2. Display will briefly read "50 mOsm Calibration" and then
prompt the user to insert a 50 mOsm calibration standard.
19

Table of Contents

Related product manuals