Model 3250/4250 Service Manual
Proper alignment and adjustment of the man-
drel, probe and stir/freeze wire are of utmost
importance to assure accuracy, precision and
proper operation of the instrument.
Note: After making any probe and/or
stir/freeze wire adjustments, the
instrument calibration must be checked
and adjusted as necessary.
Use the probe alignment tool and follow the
Probe Alignment Tool instructions located in
the Replacement Instructions section of this
manual.
Probe & Stir/Freeze Wire Alignment