Section II - Product Use
27
9-1. Situation requiring a defibrillation shock
Voice prompt: Performing an electric shock.
(Charge Beep)
Lamps to indicate the operation state: electrode
pads ( ), electrocardiogram ( ) lamps
are lit.
9-1-1. After the arrangement of electric shock energy is complete
Voice prompt: Press the flashing red button
Lamp state: Electrode pad attachment lamp ( ), and ECG analysis
lamp ( ) are ON, and electric shock button ( ) flickers.
9-1-2. Delivering electric shock energy
When electric shock button lamp ( ) flicks, the electric shock energy will be
delivered to the patient if you press the electric shock button.
Voice prompt: Electric shock is done.
Lamp state: Electrode pad attachment lamp ( ) and ECG analysis
lamp ( ) will be ON.
You must keep away from the patient because
if you touch the patient before pushing the
defibrillation shock button, you might get a
defibrillation shock also.
If you do not push the defibrillation shock
button within 20 seconds, defibrillation shock
procedure will be cancelled.
If defibrillation shock procedure is cancelled,
move on to the CPR procedure.
9-2. Situation not requiring a defibrillation shock
In case that a defibrillation shock is not required, even if you push the