2241A EN 20040210
Resetting the Digitizer
In exceptional circumstances you may be prompted to reset the CR 75.0,
either by a message on the keypad or as part of a troubleshooting procedure
in this manual.
To reset the Digitizer:
1 Locate the main switch and place it in position ‘OFF’.
2 Wait 30 seconds.
3 Place the main switch in position ‘ON’.
Never reset the Digitizer to solve a plate or cassette jam. If you would do
so, the plate inside the Digitizer might get damaged. In case of a plate or
cassette jam, always follow the procedures described in Chapter 4,
‘Preventive maintenance and troubleshooting’.