44
2241A EN 20040210
Changing the destination
If an image processing station is out of operation, you can reroute the images
to another image processing station. Start as in ‘Enabling/disabling image
transmission’ on page 42:
1 In the key-operator main menu, select ‘Digitizer set-up’ via the Up and Down
keys and confirm.
The CR 75.0 will display the Digitizer set-up menu:
# If you do not press a key within 1 minute, the CR 75.0 will exit the key-
operator mode.
2 Select the Source rerouting field via the Up and Down keys.
3 Select the image processing station from which you want to redirect the
images via the Left and Right keys.
# You can only redirect the images of one image processing station.
4 Select the Target rerouting field via the Up and Down keys.
Queue management
Digitizer set-up
Date and Time
Send test image
System info
Install
Save configuration
Fast preview
: quit
: ok
: select
KEY-OPERATOR
MENU
SERVICE XXXXX
Process.station: enabled
Reroute process.station
[PPname1]
to
[PPname2]
Scan mode: standard
Acoustic signal: enabled
: quit
: ok
: select
: change
DIGITIZER
SET-UP
SERVICE XXXXX
T
arget rerouting
f
ield
Source rerouting
field