89
2241A EN 20040210
Environmental requirements
Room temperature 20°C - 30°C
Maximum temperature change 0.5°C/min.
Relative humidity 10% - 80%
Magnetic field
Less than 5 Gauss; compliant with
EN 61000-4-8, Level 5
Sunlight exposure Not to be operated in full sunlight
Physical emissions
Noise emission (sound power level according to DIN 45635 Part.27)
• During scanning max. 65 dB(A)
• Standby max. 46 dB(A)
Heat emission (at max. throughput with 35 cm x 43 cm image plates)
• Europe 0.8 kWh
• US 2730 BTU/hr
Cassette buffer capacity
10 cassettes of mixed sizes, both in input and output buffer
Performance
Throughput for high resolution 115 plates/h (size dependent)
Throughput for standard resolution 115 plates/h (size dependent)
Grayscale resolution
Data acquisition 12 bits/pixel
Output to processor 12 bits/pixel