DR Detector Calibration|Calibrating the DR Detector|11
Preconditions
• Clean the DR Detector (see the DR Detector user manual for instructions).
• Make sure that all parts of the X-Ray system that may affect imaging (e.g. the DAP meter) are in
their normal position and clean.
• Make sure the battery is fully charged before starting the calibration (battery powered DR De-
tectors only).
• The DR Detector must be powered on before starting the NX workstation.
• Allow the DR Detector to warm up before performing the calibration. During warm-up no expo-
sures should be made. Warm-up time is one hour, unless specified otherwise.
• The ID of the DR Detector must be known.
• The X-ray generator is operational.
• Make sure that there are no foreign objects in the exposed area (e.g. stitching grid).
• Move the floating tabletop to its normal position to avoid that the tabletop borders become visi-
ble on the DR Detector image.