DR Detector Calibration|Problem solving|35
Generator error
Problem The calibration of a detector for dynamic imaging is interrupted by
a generator error. The software console displays the following error
code.
RAD_RUN_TIME_LIMIT
.
Cause The remaining heat capacity of the X-ray tube is insufficient for the
next exposure.
Workaround
1. Abort the calibration procedure.
2. Reset the generator.
3. Restart the calibration when the X-ray tube is cooled down.
4. The initial uniformity check has to be repeated, but the modes for
which calibration was already performed, can be skipped.