DR Detector Calibration|Calibrating the DR Detector|27
DX-D 40C and DX-D 40G Exposure Sequence
Modality Settings
Apply these modality settings:
Table 10: Modality settings
kV 75 kV
mAs (indicative value for first
exposure level)
25 mAs
mAs (indicative value for sec-
ond exposure level)
12.5 mAs
Focus large focus
SID Use the (average) SID for normal operation.
Position If the DR Detector is used in a fixed position, use the same posi-
tion for calibration.
If the DR Detector is used in different positions, perform the cali-
bration with the short side of the DR Detector parallel to the axis
of the X-ray tube.
Grid Remove the grid.
Filtration Mount the Agfa calibration filter (Cu 1.5 mm). Check that there
is no additional filtration.
Collimation The collimation area should extend at least 1.5 cm on each side
of the DR Detector.
Note If the exposure settings used for a previous calibration are known, use these settings.
On modalities that support automatic collimation, the collimation settings are applied automatical-
ly and must be checked by the user.