DR Detector Calibration|Calibrating the DR Detector|7
Calibration Frequency
The DR Detector must be recalibrated at regular intervals. Messages will be displayed on the NX
workstation when recalibration is required.
Perform calibration when exposure conditions have changed significantly and after a major
change to the hardware or software configuration of the X-ray system.
A DR detector used for veterinary applications has a calibration interval of 12 months, regardless
of the model.
Table 1: Calibration interval
DX-D 10C, DX-D 10G, DX-D 20C, DX-D 20G 3 months
DR 18M, DR 24M
DX-D 40C, DX-D 40G, DX-D 45C, DX-D 45G, DX-D 60C, DX-D
60G
XD 10, XD⁺10, XD 14, XD⁺14, XD 17, XD⁺17
XF⁺10, XF⁺14, XF⁺17
6 months
DR 10e C, DR 14e C, DR 17e C
DR 10s, DR 14s
DX-D 25C, DX-D 30C, DX-D 35C
Fixed DR Detector (Pixium RF4343 FL)
12 months
Fixed DR Detector (4343R, Pixium RAD 4343) 3 months or 12 months, depending
on the DR Detector model.
A timely reminder is displayed on
the NX workstation.
DR 14e G, DR 17e G
XD 10, XD⁺10, XD 14, XD⁺14, XD 17, XD⁺17 (veterinary appli-
cation)
XF⁺10, XF⁺14, XF⁺17 (veterinary application)
no calibration required
Table 2: Using the DR Detector on multiple NX workstations
DR 10s, DR 14s
DX-D 10C, DX-D 10G, DX-D 20C, DX-D 20G
DX-D 25C
DX-D 30C, DX-D 35C
DX-D 40C, DX-D 40G, DX-D 45C, DX-D 45G,
DX-D 60C, DX-D 60G
XD 10, XD 10+, XD 14, XD 14+, XD 17, XD
17+
XF⁺10, XF⁺14, XF⁺17
Each NX workstation has its own set of calibration da-
ta. The regular calibration must be performed on each
of the workstations.