DR Detector Calibration|Calibrating the DR Detector|17
1. Set the mAs value to the indicative value for the exposure level.
2. Make exposures as instructed by the DR Calibration Tool.
To avoid overheating of the X-ray tube, limit the rate of exposures to two per minute.
Calibrating DR 10s and DR 14s on DX-D 100
On a DX-D 100 mobile X-ray unit that is upgraded to support DR 10s or DR 14s, an extra step is
needed to activate the DR Detector during calibration.
In the modality position drop-down list on the software console, push the icon for free exposure
with ETHERNET written on top.