DR Detector Calibration|Calibrating the DR Detector|19
On modalities that support automatic collimation, the collimation settings are applied automatical-
ly and must be checked by the user.
Ralco collimators may limit the X-ray field. To overrule the limitation, the collimator must be un-
locked using a key and the collimation must be applied manually.
Exposure Sequence
Procedure:
1. Set the exposure parameters to the indicative value for the exposure level.
2. Make exposures as instructed by the DR Calibration Tool.
Do not initiate a next exposure before the DR Calibration Tool displays the instruction.
To avoid overheating of the X-ray tube, limit the rate of exposures to two per minute.
Warning: Check if the exposure time of the exposures is less or equal to 200 ms. If you
notice longer exposure times, contact your local service organization.