DR Detector Calibration|Calibrating the DR Detector|30
Ralco collimators may limit the X-ray field. To overrule the limitation, the collimator must be un-
locked using a key and the collimation must be applied manually.
Exposure Sequence
Procedure:
1. Set the mAs value to the indicative value for the first exposure level.
2. Make two exposures.
The DR Calibration Tool may instruct to increase or decrease the mAs setting to reach the re-
quired exposure level.
3. Make 10 exposures.
4. Set the mAs value to the indicative value for the second exposure level.
5. Make two exposures.
The DR Calibration Tool may instruct to increase or decrease the mAs setting to reach the re-
quired exposure level.
6. Make one exposure.