Preconditions
Clean the DR Detector (see the DR Detector user manual for instructions).
Make sure that all parts of the X-Ray system that may affect imaging (e.g.
the DAP meter) are in their normal position and clean.
Make sure the battery is fully charged before starting the calibration
(battery powered DR Detectors only).
The DR Detector must be powered on before starting the NX workstation.
Allow the DR Detector to warm up before performing the calibration.
During warm-up no exposures should be made. Warm-up time is one hour,
unless specified otherwise.
The ID of the DR Detector must be known.
The X-ray generator is operational.
Make sure that there are no foreign objects in the exposed area (e.g.
stitching grid).
Move the floating tabletop to its normal position to avoid that the tabletop
borders become visible on the DR Detector image.
DR Detector Calibration | Calibrating the DR Detector | 13
0134I EN 20210924 1204