About Gain Calibration
Regular calibration of the DR Detector assures constant image quality. This
calibration eliminates the effect of sensitivity deviations over the panel and/or
corrects non-unifomity in the X-ray beam. The sensitivity deviations of pixels
across the DR Detector are affected by the quantity and pattern of the effective
X-ray dose on the detector. Depending on the the panel type it will also correct
dead pixels in the acquired images.
8 | DR Detector Calibration | Calibrating the DR Detector