Fixed DR Detector (4343R, Pixium RAD 4343) Exposure
Sequence
For using the fixed DR Detector, the temperature difference between
calibration and usage must be within the recommended range of +/-6°C (for
a DR Detector with CsI conversion screen) or +/-10°C (for a DR Detector with
GOS conversion screen). Check the environmental conditions and observe the
warming-up time of the DR Detector.
Topics:
• Modality Settings
• Exposure Sequence
Modality Settings
Depending on the DR Detector model, modality settings may be specified by
the DR Calibration Tool.
If the DR Calibration Tool does not specify modality settings, refer to following
table.
Table 12: Modality settings
DR Table
DR Wall Stand DX-D 300 U-arm
kV 75 kV 75 kV 75 kV
mAs (in-
dicative
value for
first ex-
posure
level)
28 mAs 32 mAs 32 mAs
Focus large focus large focus large focus
SID Use the (average) SID for normal operation.
Position - - arm angle 0°
detector angle 0°
Grid Remove the grid.
If the grid is fixed, use 32 mAs as indicative mAs-value.
Filtration Mount the Agfa calibration filter (Cu 1.5 mm).
Check that there is no additional filtration.
DR Detector Calibration | Calibrating the DR Detector | 35