DX-D 30C and DX-D 35C Exposure Sequence
Topics:
• Modality Settings
• Exposure Sequence
Modality Settings
Apply these modality settings:
Table 9: Modality settings
kV
75 kV
mAs (indicative value for
uniformity check)
50 mAs
mAs (indicative value for
calibration)
20 mAs
Focus large focus
SID Use the (average) SID for normal operation.
Position If the DR Detector is used in a fixed position, use
the same position for calibration.
If the DR Detector is used in different positions,
perform the calibration with the short side of the
DR Detector parallel to the axis of the X-ray tube.
Grid
Remove the grid.
Filtration Mount the Agfa calibration filter (Cu 1.5 mm).
Check that there is no additional filtration.