EasyManua.ls Logo

AGFA DX-D 35C - DX-D 25 C Exposure Sequence

AGFA DX-D 35C
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DX-D 25C Exposure Sequence
Topics:
Modality Settings
Exposure Sequence
Modality Settings
Apply these modality settings:
Table 8: Modality settings
kV
75 kV
mAs (indicative value for
first exposure level)
12.5 mAs
mAs (indicative value for
second exposure level)
8 mAs
mAs (indicative value for
third exposure level)
4 mAs
Focus large focus
SID Use the (average) SID for normal operation.
Position If the DR Detector is used in a fixed position, use
the same position for calibration.
If the DR Detector is used in different positions,
perform the calibration with the short side of the
DR Detector parallel to the axis of the X-ray tube.
Grid Remove the grid.
Filtration No filtration.
DR Detector Calibration | Calibrating the DR Detector | 27
0134I EN 20210924 1204

Related product manuals