XD 10, XD⁺10, XD 14, XD⁺14, XD 17 and XD⁺17 Exposure
Sequence
Topics:
• Modality Settings
• Exposure Sequence
• Partial calibration
Modality Settings
Apply these modality settings:
Table 3: Modality settings
kV
75 kV
mAs (indica-
tive value for
first exposure
level)
25 mAs (XD 10, XD⁺10)
32 mAs (XD 14, XD⁺14, XD 17, XD⁺17)
mAs (indica-
tive value for
second expo-
sure level)
12.5 mAs (XD 10, XD⁺10)
4 mAs (XD 14, XD⁺14, XD 17, XD⁺17)
Focus large focus
SID Use the (average) SID for normal operation.
Position If the DR Detector is used in a fixed position, use the same
position for calibration.
If the DR Detector is used in different positions, perform
the calibration with the short side of the DR Detector paral-
lel to the axis of the X-ray tube.