Partial calibration
For veterinary applications, a partial calibration can be performed. The
exposed area during calibration is not covering the whole detector surface. A
minimum exposed area of 10 by 12 cm is required.
WARNING:
It is not allowed to use a partially calibrated DR detector for
human patients.
Procedure:
Make the exposures as instructed by the DR Calibration Tool.
The tool will detect the partial exposure and ask for confirmation that a
partial calibration can be applied.
DR Detector Calibration | Calibrating the DR Detector | 19