Generator error
Problem The calibration of a detector for dynamic imaging is interrupted
by a generator error. The software console displays the following
error code.
RAD_RUN_TIME_LIMIT
.
Cause The remaining heat capacity of the X-ray tube is insufficient for
the next exposure.
Work-
around
1. Abort the calibration procedure.
2. Reset the generator.
3. Restart the calibration when the X-ray tube is cooled down.
4. The initial uniformity check has to be repeated, but the
modes for which calibration was already performed, can be
skipped.
DR Detector Calibration | Problem solving | 41
0134I EN 20210924 1204