DR Calibration Tool Stalls
Problem The DR Calibration Tool stalls.
Cause A virus scanner interferes with the DR Cali-
bration Tool.
Workaround Disable the virus scanners on the workstation
and restart the calibration procedure.
Problem The DR Calibration Tool stalls for a short
time.
Cause Information must be retrieved from the DR
Detector.
Workaround Wait until the communication to the DR De-
tector has finished and the DR Calibration
Tool reacts again. This may take a number of
seconds.
40 | DR Detector Calibration | Problem solving
0134I EN 20210924 1204