DX-D 40C and DX-D 40G Exposure Sequence
Topics:
Modality Settings
Exposure Sequence
Modality Settings
Apply these modality settings:
Table 10: Modality settings
kV
75 kV
mAs (indicative value for
first exposure level)
25 mAs
mAs (indicative value for
second exposure level)
12.5 mAs
Focus large focus
SID Use the (average) SID for normal operation.
Position If the DR Detector is used in a fixed position, use
the same position for calibration.
If the DR Detector is used in different positions,
perform the calibration with the short side of the
DR Detector parallel to the axis of the X-ray tube.
Grid Remove the grid.
Filtration Mount the Agfa calibration filter (Cu 1.5 mm).
Check that there is no additional filtration.
DR Detector Calibration | Calibrating the DR Detector | 31