45
2321 E EN 20130722
DX-G/DX-M
The digitizer complies with the EN 60601-1 and UL 60601-1 standards for Med-
ical Electrical Equipment. This means that, although it is absolutely safe,
patients may not come in direct contact with the equipment. Therefore the
operator console must be placed outside the radius as listed below around the
patient (according to the local valid regulation).
Figure 1: Patient environment
R = 1.5 m/4.9 feet (EN 60601-1) or 1.83 m/6 feet (UL 60601-1).
h = 2.5 m/8.2 feet (EN 60601-1) or 2.29 m/7.5 feet (UL 60601-1).
Perform no other operations on the digitizer than those described in this docu-
ment.
Switch the system off before moving it. When reaching the new position, fix the
digitizer in place and switch the system on again.
The digitizer is suitable for use in all establishments other than domestic, and
may be used in domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings used for
domestic purposes, provided the warning concerning radio interference in the
paragraph ‘Safety Directions’ is heeded.
Patient vicinity
Patient vicinity
R
h