92
2321 E EN 20130722
DX-G/DX-M
The distance can be determined through the equation for each respective column.
P is the rated power of the transmitter in watts (W) according to the manufacturer infor-
mation on the transmitter, only for transmitters where the rated power is not mentioned
in the above table.
REMARK 1 : An additional factor of 10/3 has been used to calculate the recommended
protective distance of transmitters in the frequency range from 80 MHz to 2.5 GHz, to
reduce the probability that mobile portable communication equipment unintentionally
brought into the area of the patients will lead to a disruption.
REMARK 2 : These Guidelines may not be relevant in all situations. The dispersion of
electromagnetic waves is influenced by absorption and reflections from buildings,
objects and people.