2 Site Requirements for the Agilent 2100 Bioanalyzer
Environment
Environment
Your Agilent 2100 Bioanalyzer will work within the specifications of ambient
temperatures and relative humidity described in Table 1 on page 20.
CAUTION
Condensation
Condensation will damage the electronics.
Do not store, ship or use the instrument under conditions where temperature
fluctuations could cause condensation within the instrument.
If your instrument was shipped in cold weather, leave it in its box and allow it to
warm up slowly to room temperature to avoid condensation.
NOTE
The Agilent 2100 Bioanalyzer is designed to operate in a controlled electromagnetic
environment (EN61326/A1) where RF transmitters such as mobile telephones should not
be used in close proximity.
NOTE
In order to function properly, do not place the Agilent 2100 Bioanalyzer on a vibrating
surface or near vibrating objects.