2 Site Requirements for the Agilent 2100 Bioanalyzer
Physical Specifications of the Agilent 2100 Bioanalyzer
Physical Specifications of the Agilent 2100 Bioanalyzer
Table 1 on page 20 lists the physical specifications of the Agilent 2100
Bioanalyzer.
Tab le 1 Physical specifications
Type Specification Comment
Weight 10 kg (22 lbs)
Dimensions 162 — 412 — 290 mm
(6.4 — 16.2 — 11.4 in)
Width — depth — height
Line voltage 100-240 VAC ±10% Wide-ranging capability
Line frequency 50-60 Hz ±5%
Power consumption 30 W / 60 VA Maximum
Ambient operating temperature 5-40 °C (41-104 °F) The assays set-up to be analyzed on a chip
temperature of 30 °C. If the ambient
temperature is above 30 °C the assays are
analyzed using that temperature.
Ambient non-operating
temperature
-40-70 °C (-40-158 °F)
Humidity <80%, at 5-31 °C (41-89 °F), decreasing
linearly to 50% at 40 °C (104 °F)
Non-condensing
Operating altitude Up to 2000 m (6500 ft)
Non-operating altitude Up to 4600 m (14950 ft) For storage of the Agilent 2100 Bioanalyzer
Safety standards: IEC, CSA, UL Installation Category II, Pollution
Degree 2