EasyManua.ls Logo

Agilent Technologies 2100 - Physical Specifications of the Agilent 2100 Bioanalyzer

Agilent Technologies 2100
42 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
20
2 Site Requirements for the Agilent 2100 Bioanalyzer
Physical Specifications of the Agilent 2100 Bioanalyzer
Physical Specifications of the Agilent 2100 Bioanalyzer
Table 1 on page 20 lists the physical specifications of the Agilent 2100
Bioanalyzer.
Tab le 1 Physical specifications
Type Specification Comment
Weight 10 kg (22 lbs)
Dimensions 162 — 412 — 290 mm
(6.4 — 16.2 — 11.4 in)
Width — depth — height
Line voltage 100-240 VAC ±10% Wide-ranging capability
Line frequency 50-60 Hz ±5%
Power consumption 30 W / 60 VA Maximum
Ambient operating temperature 5-40 °C (41-104 °F) The assays set-up to be analyzed on a chip
temperature of 30 °C. If the ambient
temperature is above 30 °C the assays are
analyzed using that temperature.
Ambient non-operating
temperature
-40-70 °C (-40-158 °F)
Humidity <80%, at 5-31 °C (41-89 °F), decreasing
linearly to 50% at 40 °C (104 °F)
Non-condensing
Operating altitude Up to 2000 m (6500 ft)
Non-operating altitude Up to 4600 m (14950 ft) For storage of the Agilent 2100 Bioanalyzer
Safety standards: IEC, CSA, UL Installation Category II, Pollution
Degree 2

Other manuals for Agilent Technologies 2100

Related product manuals