Chapter 4 Remote Interface Reference
Frequency Sweep Commands
4
Sweep Commands
FREQuency:STARt {<frequency>|MINimum|MAXimum}
FREQuency:STARt? [MINimum|MAXimum]
Set the start frequency (used in conjunction with the stop frequency).
Select from 1
µ
Hz to 20 MHz
(limited to 200 kHz for ramps and 6 MHz
for arbitrary waveforms). The default is 100 Hz. MIN = 1 µHz. MAX =
20 MHz. The :STAR? query returns the start frequency in hertz.
•To sweep up in frequency, set the start frequency < stop frequency.
To sweep
down
in frequency, set the start frequency
>
stop frequency.
FREQuency:STOP {<frequency>|MINimum|MAXimum}
FREQuency:STOP? [MINimum|MAXimum]
Set the stop frequency (used in conjunction with the start frequency).
Select from 1
µ
Hz to 20 MHz
(limited to 200 kHz for ramps and 6 MHz
for arbitrary waveforms). The default is 1 kHz. MIN = 1 µHz. MAX =
20 MHz. The :STOP? query returns the stop frequency in hertz.