Chapter 3 Features and Functions
Pulse Width Modulation (PWM)
3
Pulse Waveform
Pulse is the only waveform shape supported for PWM.
Front-Panel Operation:Press .
Remote Interface Operation:
FUNCtion {PULSe}
You can also use the
APPLy
command to select the function, frequency
,
amplitude, and offset with a single command.
Pulse Period
The range for the pulse period is 200 ns to 2000 s. The default is 1 ms.
Front-Panel Operation: After selecting the pulse function, press the
Freq softkey again to toggle to the Period softkey. Then use the knob
or numeric keypad to enter the desired pulse period.
Remote Interface Operation:
PULSe:PERiod {<seconds>|MINimum|MAXimum}