303
Chapter 5 Error Messages
Device Dependent Errors
4
5
Device Dependent Errors
-313 Calibration memory lost;
memory corruption detected
The non-volatile memory used to store the function generator’s calibration
constants has detected a checksum error. This error may be the result of
a device failure or extreme conditions such as lightning or strong
magnetic fields.
-314 Save/recall memory lost;
memory corruption detected
The non-volatile memory used to store instrument states has detected a
checksum error. This error may be the result of a device failure or
extreme conditions such as lightning or strong magnetic fields.
-315 Configuration memory lost;
memory corruption detected
The non-volatile memory used to store the function generator’s
configuration settings (i.e., remote interface settings) has detected a
checksum error. This error may be the result of a device failure or
extreme conditions such as lightning or strong magnetic fields.
-350 Queue overflow
The error queue is full because more than 20 errors have occurred.
No additional errors are stored until you remove errors from the queue.
The error queue is cleared by the *CLS (clear status) command or when
the power is cycled. The errors are also cleared when you read
the queue.
The error queue
is not
cleared by an instrument reset
(*RST command).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33220A and is the answer not in the manual?

Agilent Technologies 33220A Specifications

General IconGeneral
BrandAgilent Technologies
Model33220A
CategoryInverter
LanguageEnglish

Related product manuals