EasyManuals Logo
Home>Agilent Technologies>Inverter>33220A

Agilent Technologies 33220A User Manual

Agilent Technologies 33220A
363 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #273 background imageLoading...
Page #273 background image
272
Chapter 4 Remote Interface Reference
Status Reporting Commands
4
See the table on page 268 for the register bit definitions.
*ESR?
Query the Standard Event Status Register
. Once a bit is set, it remains
set until cleared by a *CLS (clear status) command or queried by this
command. A query of this register returns a decimal value which
corresponds to the binary-weighted sum of all bits set in the register.
*ESE <enable value>
*ESE?
Enable bits in the Standard Event Status Register to be reported in the
Status Byte. The selected bits are summarized in the “Standard Event” bit
(bit 5) of the Status Byte Register. The *ESE? query returns a decimal
value which corresponds to the binary-weighted sum of all bits enabled
by the *ESE command.
•A *CLS (clear status) does not clear the enable register but it does
clear all bits in the event register.
•A STATus:PRESet does not clear the bits in the Status Byte
enable register.
•A *PSC 0 preserves the contents of the enable register through
power cycles.
Standard Event Register Commands

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33220A and is the answer not in the manual?

Agilent Technologies 33220A Specifications

General IconGeneral
BrandAgilent Technologies
Model33220A
CategoryInverter
LanguageEnglish

Related product manuals