Chapter 4 Remote Interface Reference
Interface Configuration Commands
4
SYSTem:COMMunicate:LAN:LIPaddress?
Returns the “last valid Auto IP address used” in dot notation (for
example: “169.254.20.11”).
SYSTem:COMMunicate:LAN:MAC?
Returns the 33220A’s Media Access Control (MAC) address. This is an
unchangeable 48-bit address assigned by the manufacturer to each
unique internet device. The MAC address is represented as 12
hexadecimal characters in the form “XX-XX-XX-XX-XX-XX” (for
example, “00-30-D3-00-10-41”).
SYSTem:COMMunicate:LAN:MEDiasense {OFF|0|ON|1}
SYSTem:COMMunicate:LAN:MEDiasense?
Disables or enables the MEDiasense feature. When enabled, this feature
detects a loss of LAN connectivity greater than 20 seconds, and then, on
sensing a reconnection, re-starts the LAN. The factory setting is enabled
(ON). The query returns “0” (OFF) or “1” (ON).
If you change the MEDiasense setting, you must cycle power on the
instrument to activate the new setting.
The MEDiasense setting is stored in non-volatile memory, and does
not change when power has been off, or after a instrument reset
(*RST).
SYSTem:COMMunicate:LAN:NETBios {OFF|0|ON|1}
SYSTem:COMMunicate:LAN:NETBios?
Disables or enables the use of NETBios to automatically assign an IP
address to the 33220A for peer-to-peer naming on a private network that
uses the NETBios naming service. The factory setting is enabled (ON).
The query returns “0” (OFF) or “1” (ON).
If you change the NETBios setting, you must cycle power on the
instrument to activate the new setting.
The NETBios setting is stored in non-volatile memory, and does not
change when power has been off, or after a instrument reset (*RST).