98
Chapter 3 Features and Functions
Pulse Width Modulation (PWM)
3
Modulating Source
The function generator will accept an internal or external modulation
source for PWM.
Modulating source: Internal or External. The default is Internal.
If you select the External source, the pulse waveform is modulated
with an external waveform. The width deviation or duty cycle
deviation is controlled by the ±5V signal level present on the rear-
panel Modulation In connector. For example, if you have set the duty
cycle to 10% and the duty cycle deviation to 5%,
then when the
modulating signal is at +5 volts, the output will be at the
maximum
duty cycle (15%). When the modulating signal is at -5 volts, the output
will be at the
minimum
duty cycle (5%).
Front-Panel Operation
: After enabling PWM, press the
Source
softkey.
Remote Interface Operation
:
PWM:SOURce {INTernal|EXTernal}
Modulation
In
+5 V
0V
-5 V

Table of Contents