Agilent 4UHV Ion Pump Controller User Manual / 87-900-137-01 (A) 47/104
Installation
WARNING!
For the safety of the operator, the 4UHV controller must be powered with a 3-
wire power cable connected to a plug approved at international level. Use
this cable and plug together with an adequately earthed socket so as to
prevent electrical shocks and satisfy the requirements of EC norms. The high
voltages that develop in the controller can provoke serious injuries or death.
After switching off the unit, residual energy is present inside the controller for
some time.
Wait approx. 1 minute to be sure that the residual energy has been
dissipated.
CAUTION!
The controller can be used either as a desktop unit or as a rack module. In any
case it must be positioned in a way that the air can circulate freely through the
ventilation slots on the cover.
If the controller is used as a rack module, it MUST be inserted into a four-rack
high adapter to prevent it from falling inside the rack itself (see the table of
orderable parts). The 4UHV controller front panel is not designed to support the
weight of the unit.
Do not install or use the controller in an environment exposed to atmospheric
agents (rain, snow, ice), in the presence of dust, corrosive gases or in a highly
flammable or explosive environment.