Technical Information
Specifications
5
Agilent 4UHV Ion Pump Controller User Manual / 87-900-137-01 (A) 59/104
Specifications
Table 1 Specifications
Component Description
Input Voltage 100 - 240 Vac (+/-10%)
Input Frequency 50/60 Hz
Input Power 600 VA
Output HV Voltage +/- 7000 Vdc +/- 5%
Output HV Short
circuit current
40 mA (for 80W channels) 100 mA (for 200 W channels)
Output HV Power
max
200 W or 80 W each channel
Operating
temperature
0 °C to 40 °C
Storage
temperature
-20 °C to +70 °C
Fuse 5x20 T6.3A 250V
Voltage
measurement
Resolution 100 V
Accuracy: +/- 5%
Current
measurement
Measurement range: 10 nA to 100 mA, logarithmic scale
Resolution : 2 digit of the mantissa (maximum 10 nA)
Accuracy: +/- 6%
Weight 6.5 kg
Compliance with: EN 61010-1 2001: Safety requirements for electrical
equipment for measurement control and laboratory use.
EN 61326-1 2006: Electrical equipment for measurement
control and laboratory use EMC requirements
This Manual: http://www.manuallib.com/file/2630540