EasyManuals Logo

Agilent Technologies 4UHV User Manual

Agilent Technologies 4UHV
105 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #60 background imageLoading...
Page #60 background image
Technical Information
Specifications
5
Agilent 4UHV Ion Pump Controller User Manual / 87-900-137-01 (A) 59/104
Specifications
Table 1 Specifications
Component Description
Input Voltage 100 - 240 Vac (+/-10%)
Input Frequency 50/60 Hz
Input Power 600 VA
Output HV Voltage +/- 7000 Vdc +/- 5%
Output HV Short
circuit current
40 mA (for 80W channels) 100 mA (for 200 W channels)
Output HV Power
max
200 W or 80 W each channel
Operating
temperature
0 °C to 40 °C
Storage
temperature
-20 °C to +70 °C
Fuse 5x20 T6.3A 250V
Voltage
measurement
Resolution 100 V
Accuracy: +/- 5%
Current
measurement
Measurement range: 10 nA to 100 mA, logarithmic scale
Resolution : 2 digit of the mantissa (maximum 10 nA)
Accuracy: +/- 6%
Weight 6.5 kg
Compliance with: EN 61010-1 2001: Safety requirements for electrical
equipment for measurement control and laboratory use.
EN 61326-1 2006: Electrical equipment for measurement
control and laboratory use EMC requirements
This Manual: http://www.manuallib.com/file/2630540

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4UHV and is the answer not in the manual?

Agilent Technologies 4UHV Specifications

General IconGeneral
BrandAgilent Technologies
Model4UHV
CategoryController
LanguageEnglish

Related product manuals