Installation
42 Agilent 5100 and 5110 ICP-OES User’s Guide
Argon Humidifier
The Argon Humidifier is commonly used when running aqueous
samples with high dissolved solids or high dissolved salt content.
When using the accessory, the nebulizer gas flow is passed through
the humidifier to increase the water vapor in the gas. This has been
found to be beneficial by reducing the build-up of salt and other
dissolved solids in the sample introduction system. By reducing
blockages, the Argon Humidifier helps to ensure uninterrupted,
maintenance-free operation.
IsoMist Programmable Temperature Spray Chamber
IsoMist utilizes the Peltier effect to provide electrical cooling or
heating to the spray chamber.
The ‘Set Temperature’ is monitored and controlled via the ICP Expert
software. The IsoMist performance can be monitored and controlled
from a lab PC via USB or Bluetooth wireless technology.