EasyManuals Logo

Agilent Technologies 5977C Inert+ MSD User Manual

Agilent Technologies 5977C Inert+ MSD
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #136 background imageLoading...
Page #136 background image
5 Operating in CI Mode
To Verify NCI Performance
136 5977C GC/MSD Operating Manual
To Verify NCI Performance
This procedure is for EI/PCI/NCI MSDs only.
Materials needed
OFN, 100 fg/µL 3 × 1 mL (5188-5347)
Procedure
1 Verify that the MSD performs correctly in EI mode.
2 Load the NCICH4.U tune file, and accept the temperature setpoints.
3 Select Gas A and set flow to 40%.
4 In Tune and Vacuum Control view, run CI Autotune. (See “To Perform an NCI
Autotune (Methane Reagent Gas)” on page 133.)
Note that there are no criteria for a “passing” Autotune in CI. If the Autotune
completes, it passes.
5 Run the NCI sensitivity method: OFN_NCI.M using 2 µL of 100 fg/µL OFN.
6 Verify that the system conforms to the published sensitivity specification.
Please refer to the Agilent Web site at www.agilent.com/chem for
specifications.
CAUTION
Always verify MSD performance in EI before switching to CI operation. (See
“To Verify EI System Performance” on page 104.) Always set up the CI MSD in
PCI first, even if you are going to run NCI.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977C Inert+ MSD and is the answer not in the manual?

Agilent Technologies 5977C Inert+ MSD Specifications

General IconGeneral
BrandAgilent Technologies
Model5977C Inert+ MSD
CategorySecurity Sensors
LanguageEnglish

Related product manuals