2 Installing 8890 GC Columns
To Install a Capillary Column in the GC/MSD Interface Using the Self-Tightening Column Nut
42 5977C GC/MSD Operating Manual
To Install a Capillary Column in the GC/MSD
Interface Using the Self-Tightening Column
Nut
This procedure is for the installation of a capillary column directly into the
analyzer using the Agilent recommended self-tightening column nut.
Materials needed
Interface tip seal (G3870-20542) (See Figure 6 on page 44.) (Not used with EI
SS or Inert source)
Knurled tip seal retainer (G3870-20547) (Not used with EI SS or Inert source)
Tip seal spring (G7005-20024) (Not used with EI SS or Inert source)
Column cutter, ceramic (5181-8836) or diamond (5183-4620)
Flashlight
Magnifying loupe
Gloves, clean
Large (8650-0030)
Small (8650-0029)
Self Tightening column nut for GC\MSD interface (5190-5233)
Ferrules, Vespel
0.27 mm id, for 0.10 mm id columns (5062-3518)
0.37 mm id, for 0.20 mm id columns (5062-3516)
0.40 mm id, for 0.25 mm id columns (5181-3323)
0.5 mm id, for 0.32 mm id columns (5062-3514)
0.8 mm id, for 0.53 mm id columns (5062-3512)
Septum (may be old, used inlet septum)
Safety glasses

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977C Inert+ MSD and is the answer not in the manual?

Agilent Technologies 5977C Inert+ MSD Specifications

General IconGeneral
BrandAgilent Technologies
Model5977C Inert+ MSD
CategorySecurity Sensors
LanguageEnglish

Related product manuals