7CI Maintenance
To Switch from the EI HES to the CI Source
5977C GC/MSD Operating Manual 205
To Switch from the EI HES to the CI Source
Procedure
1 Vent the MSD. (See“To Vent the MSD” on page 109.) The software will
prompt you for the appropriate actions.
2 Open the left side access panel. (See“To Open the MSD Covers” on
page 108.)
3 Open the analyzer chamber. (See“To Open the Analyzer Chamber” on
page 149.)
4 Remove the EI HES. (See“To Remove the EI HES” on page 151.)
5 Make sure the EI HES source has cooled and then place it in its storage
container.
6 Remove the EI HES radiator. (See“To Remove the EI HES Radiator” on
page 206.)
7 Place the EI HES radiator in the storage container.
8 Remove the CI source radiator from the storage container.
9 Install the CI source radiator. (See“To Install the CI Source Radiator” on
page 209.)
10 Remove the CI source from the storage container.
11 Install the CI source. This requires shortening the column so that it protrudes
from the transfer line by 1 to 2 mm. (See“To Install the CI Source” on
page 232.)
12 Pumpdown the MSD. (See“To Pumpdown the MSD in CI Mode” on
page 120.)
CAUTION
Always verify GC/MSD performance in EI before switching to CI operation.
CAUTION
Electrostatic discharges to analyzer components are conducted to the side
board where they can damage sensitive components. Wear a grounded
antistatic wrist strap. (See“Electrostatic discharge is a threat to MSD
electronics” on page 20.) Take antistatic precautions before you open the
analyzer chamber.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977C Inert+ MSD and is the answer not in the manual?

Agilent Technologies 5977C Inert+ MSD Specifications

General IconGeneral
BrandAgilent Technologies
Model5977C Inert+ MSD
CategorySecurity Sensors
LanguageEnglish

Related product manuals