3 Installing Intuvo 9000 GC Columns
To Replace the 9000 GC/MS Tail
68 5977C GC/MSD Operating Manual
To Replace the 9000 GC/MS Tail
This procedure is necessary if you are switching to a source that requires a
different 9000 GC/MS Tail, replacing a leaking gasket or a contaminated
9000 GC/MS Tail, or separating the 9000 GC from the MSD.
Materials needed
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Tweezers (8710-2460)
9000 GC/MS Tail, used with a standard source (G4590-60009)
9000 GC/MS Tail, used with a HES (G4590-60109)
Intuvo gasket, polyimide 5/pk, for temperatures < 350 °C (5190-9072)
Intuvo gasket, nickel 5/pk, for temperatures 350 °C to 450 °C (5190-9073)
Intuvo torque driver stored on oven door (5190-9571)
7/16-inch open-end wrench
Procedure
1 Vent the MSD (See “To Vent the MSD” on page 109.) When prompted, set
the temperature for the ion source, quads, inlet, detector, column, Guard Chip,
9000 GC/MS Tail, and other components in the flow path heated zones to
<40 °C.
2 If you are using hydrogen or other flammable gas as a carrier gas, close the
manual gas supply valve to the instrument before turning off the MSD power.
3 Wait until the GC reaches the ready state indicating the components are
cooled below 40 °C before continuing the steps in this procedure.
WARNING
The inlet, detector, bus components, and column can be hot enough to cause
burns. Cool heated zones to a safe handling temperature before continuing.
CAUTION
Always wear clean gloves while handling any parts that are in the flow path
going to the analyzer or parts that are located inside the analyzer.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977C Inert+ MSD and is the answer not in the manual?

Agilent Technologies 5977C Inert+ MSD Specifications

General IconGeneral
BrandAgilent Technologies
Model5977C Inert+ MSD
CategorySecurity Sensors
LanguageEnglish

Related product manuals