Page #175 background image
6 General Maintenance
To Disassemble the EI XTR Source
5977C GC/MSD Operating Manual 175
To Disassemble the EI XTR Source
Materials needed
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Hex ball driver, 1.5 mm (8710-1570)
Hex ball driver, 2.0 mm (8710-1804)
Wrench, open-end, 10 mm (8710-2353)
Procedure
Refer to the exploded parts view Figure 57 and the EI EXT parts list Table 20 on
page 176 while using this procedure.
1 Remove the EI XTR source. (See “To Remove the EI XTR, SS, Inert, or
Hydronert Source” on page 169.)
2 Remove the filaments by removing the two gold plated screws and
separating the filaments from the source.
3 Loosen the two gold plated screws on the source heater block assembly, and
separate the repeller assembly from the source body. The repeller assembly
includes the source heater block assembly, repeller, and related parts.
4 Remove the gold plated setscrews from the side of the source body.
5 Pull the entrance lens and ion focus lens to remove them from the source
body.
6 Remove the extractor lens and insulator.
7 Separate the entrance lens and ion focus lens from the lens insulator.
8 Remove the repeller nut, washers, and insulator from the front side of the
source heater block assembly, then remove the repeller, insulator, and
repeller block insert from the opposite side.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977C Inert+ MSD and is the answer not in the manual?

Agilent Technologies 5977C Inert+ MSD Specifications

General IconGeneral
BrandAgilent Technologies
Model5977C Inert+ MSD
CategorySecurity Sensors
LanguageEnglish

Related product manuals