6 General Maintenance
Maintaining the Vacuum System
146 5977C GC/MSD Operating Manual
Maintaining the Vacuum System
Periodic maintenance
Some maintenance tasks for the vacuum system must be performed
periodically. (See Table 16 on page 141.) These include:
Checking the foreline pump fluid (every week)
Replacing the tip seal on the optional IDP-3 dry scroll pump (annually)
Checking for leaks (every month)
Checking the calibration vial(s) (every 6 months)
Replacing the foreline pump oil (every 6 months; every 3 months for CI MSDs
using ammonia reagent gas)
Tightening the foreline pump oil box screws (first oil change after installation)
Replacing the diffusion pump fluid (once a year)
Replacing the dry foreline pump seals (once a year)
Failure to perform these tasks as scheduled can result in decreased instrument
performance. It can also result in damage to your instrument.
Other procedures
Tasks such as replacing a foreline vacuum gauge or Micro-Ion vacuum gauge
should be performed only when needed. Refer to the Agilent 5977C Series MSD
Troubleshooting and Maintenance manual and the online help in Agilent
MassHunter GC/MS Acquisition software for symptoms that indicate this type of
maintenance is required.
More information is available
If you need more information about the locations or functions of vacuum system
components, refer to the Agilent 5977C Series MSD Troubleshooting and
Maintenance manual.
Most of the procedures in this chapter are illustrated with video clips. See
“Where to Find More Information” on page 4 for details.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977C Inert+ MSD and is the answer not in the manual?

Agilent Technologies 5977C Inert+ MSD Specifications

General IconGeneral
BrandAgilent Technologies
Model5977C Inert+ MSD
CategorySecurity Sensors
LanguageEnglish

Related product manuals