7CI Maintenance
To Clean the CI Source
5977C GC/MSD Operating Manual 225
5 Ultrasonically clean the parts (each group separately) for 15 minutes. For
dirty parts, use all three solvents in the order shown, cleaning 15 minutes with
each of the following solvents:
Methylene chloride (reagent-grade)
Acetone (reagent-grade)
Methanol (reagent-grade)
For routine cleaning, cleaning with methanol is sufficient.
6 Place the parts in a clean beaker. Loosely cover the beaker with clean
aluminum foil (dull side down).
7 Dry the cleaned parts in an oven at 100 °C for five to six minutes.
WARNING
Al
All of these solvents are hazardous. Work in a fume hood and take all
appropriate precautions.
WARNING
Let
Let the parts cool before you handle them.
NOTE
Take care to avoid contaminating cleaned and dried parts. Put on new, clean
gloves before handling the parts. Do not set the cleaned parts on a dirty
surface. Set them only on clean, lint-free cloths.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977C Inert+ MSD and is the answer not in the manual?

Agilent Technologies 5977C Inert+ MSD Specifications

General IconGeneral
BrandAgilent Technologies
Model5977C Inert+ MSD
CategorySecurity Sensors
LanguageEnglish

Related product manuals