EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>6430

Agilent Technologies 6430 User Manual

Agilent Technologies 6430
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #27 background imageLoading...
Page #27 background image
Inner Workings – Triple Quadrupole MS versus Single Quadrupole MS 2
Innovative Enhancements in the 6430 Triple Quadrupole
Agilent 6400 Series Triple Quad LC/MS Concepts Guide 27
Innovative Enhancements in the 6430 Triple Quadrupole
The 6430 Triple Quadrupole has many of the innovative
enhancements that are included in the Agilent 6460 Triple
Quadrupole.
The same resistive capillary transfer tube (denoted as 1 on
the 6430 in
Figure 7 on page 22)
Improved pumping in vacuum stage 2 (denoted as 2 on the
6430 in
Figure 7)
A second turbo pump has been added (denoted as 3 on the
6430 in
Figure 7)
Innovative Enhancements in the 6420 Triple Quadrupole
The 6420 Triple Quadrupole has some of the innovative
enhancements that are included in the Agilent 6430 Triple
Quadrupole.
The same resistive capillary transfer tube (denoted as 1 on
the 6420 in
Figure 7 on page 22)
The 6420 Triple Quadrupole can be upgraded to a 6430 and
then to a 6460 Triple Quadrupole. (denoted as 2 on the 6420
in
Figure 7 on page 22)
How a triple quadrupole mass spectrometer works
Quadrupoles provide a user with the capability to do MS/MS in
several ways (see Figure 12).

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6430 and is the answer not in the manual?

Agilent Technologies 6430 Specifications

General IconGeneral
BrandAgilent Technologies
Model6430
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals