EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>6430

Agilent Technologies 6430 User Manual

Agilent Technologies 6430
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #67 background imageLoading...
Page #67 background image
Agilent MassHunter Workstation Software - Data Acquisition for 6400 Series Triple Quadrupole 4
Tuning
Agilent 6400 Series Triple Quad LC/MS Concepts Guide 67
Tuning
Autotune and Checktune
A Checktune can be used to determine if the tuning mix ion
masses are properly assigned and if the response or sensitivity
of these ions is within expectations. This check tune takes up to
15 minutes to run. An Autotune can be used if a more extensive
tune is recommended (Figure 35). For all models except the
6490, an Autotune can take 15 to 20 minutes for each polarity;
the 6490 takes approximately 45 minutes per ionization mode.
Everything is automatic since the tuning mix is delivered by the
calibrant delivery system (CDS), which is switched on
automatically during the tune.
Figure 35 Autotune in progress

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6430 and is the answer not in the manual?

Agilent Technologies 6430 Specifications

General IconGeneral
BrandAgilent Technologies
Model6430
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals